-40%
Keysight Agilent HP Hewlett-Packard 34970A Data Acquisition / Switch Unit
$ 422.37
- Description
- Size Guide
Description
Keysight Agilent HP Hewlett-Packard 34970A Data Acquisition / Switch Unit.Features
3-slot mainframe with built-in GPIB and RS232 interfaces
6 1/2-digit (22-bit) internal DMM, scanning up to 250 channels per second
8 switch and control plug-in modules to choose from
Built-in signal conditioning measures thermocouples, RTDs and thermistors, AC/DC volts and current; resistance; frequency and period
50k readings of non-volatile memory holds data when power is removed
Hi/LO alarm limits on each channel, plus 4 TTL alarm outputs
A BenchVue software license (BV0006B) is now included with your instrument. BenchVue makes it simple to connect, control instruments, and automate test sequences
The Keysight 34970A data acquisition / data logger switch consists of a three-slot mainframe with a built-in 6 1/2-digit digital multimeter. You can configure each channel independently to measure one of 11 different functions without the added cost or hassles of signal-conditioning accessories. Choose from eight optional plug-in modules to create a compact data logger, full-featured data acquisition system or low-cost switching unit. On-module screw-terminal connections eliminate the need for terminal blocks and a unique relay maintenance feature counts every closure on every switch for easy, predictable relay maintenance. Our most popular module 34901A features a built-in thermocouple reference and 20 two-wire channels.
Keysight’s BenchVue DAQ Control and Analysis app is available for use with the DAQ970A, 34970A, 34972A and 34980A. BenchVue gives you an intuitive interface for test configuration and real-time data display and analysis. Setup and make measurements quickly, export data or log and view real-time measurements with a variety of display types to interpret your data faster. BenchVue Software also supports many other Keysight instruments giving you a single interface for Keysight’s most popular instruments.
Microsoft and Windows are U.S. registered trademark of Microsoft Corporation.
Counterfeits of this or closely related products have been found in the market. To ensure you are purchasing an authentic Keysight product with the quality and performance you expect, purchase through Keysight authorized distribution partners. Click on the “How to Buy or Rent” button, above, to find authorized distribution partners.
The 34970A will be discontinued on June 1, 2021. The last day to place order for this product is May 31, 2021. Keysight will continue to provide world-class support for this product for the standard period of 5 years.
34970A/34972A accuracy specifications ± (% of reading + % of range) 1
Includes measurement error, switching error, and transducer conversion error
Range 3
Frequency, etc.
24 hour 2 23 ± 1 °C
90 Day 23 ± 5 °C
1 Year 23 ± 5 °C
Temperature coefficient 0 – 18 °C, 28 - 55 °C
DC voltage
True RMS AC voltage 4
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
300.000 V
All ranges from 100.0000 mV to 100.0000 V
300.0000 V
100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ 100 mV to 300V
0.0030 + 0.0035
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0020 1.00 + 0.03
0.35 + 0.03
0.04 + 0.03
0.10 + 0.05
0.55 + 0.08
4.00 + 0.50
1.00 + 0.05
0.35 + 0.05
0.04 + 0.05
0.10 + 0.10
0.55 + 0.20
4.00 + 1.25 0.0030 + 0.0035
0.0020 + 0.0006
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010 0.10
0.05
0.03
0.006
0.005 + 0.010
0.010 + 0.004
0.050 + 0.006 1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.00 + 0.5
0.30 + 0.5
0.10+ 0.5 1.2°C
1.0°C
1.0°C
1.0°C
1.0°C
1.2°C
1.2°C
1.0°C
0.06 °C
0.08 °C
1 Adc and ac
0.0040 + 0.0040
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0030 1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.08
0.35 + 0.08
0.05 + 0.08
0.11 + 0.12
0.60 + 0.20
4.00 + 1.25 0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.010 0.10
0.05
0.03
0.01
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010 1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.00 + 0.5
0.30 + 0.5
0.10 + 0.5
400 to 1100 °C
–200 to -150 °C
–210 to -150 °C
–200 to -100 °C
–200 to -100 °C
–50 to 300 °C
–50 to 400 °C
–200 to -100 °C
0.0050 + 0.0040
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0030 1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.08
0.35 + 0.08
0.06 + 0.08
0.12 + 0.12
0.60 + 0.20
4.00 + 1.25 0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.010 0.10
0.05
0.03
0.01
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010 1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.00 + 0.5
0.30 + 0.5
0.10+ 0.5 1.8°C
1.5°C
1.2°C
1.5°C
1.5°C
1.8°C
1.8°C
1.5°C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003 0.100 + 0.004
0.035 + 0.004
0.005 + 0.004
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02
0.100 + 0.008
0.035 + 0.008
0.005 + 0.008
0.011 + 0.012
0.060 + 0.020
0.20 + 0.05 0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002 0.005
0.005
0.001
0.001 0.002
0.002
0.005 0.100
0.035
0.015
0.100
0.035
Resistance 6
Frequency and period 7
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
100 kHz – 300 kHz 5
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
100 kHz – 300 kHz 5 1 mA current source
1 mA
100μA
10μA
5.0 μA
500 nA
500 nA || 10 MΩ 3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 40 Hz
40 Hz – 300 kHz < 0.1 V burden
< 0.6 V
<2 V
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 5 kHz
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 5 kHz 1100 to 1820 °C
–150 to 1000 °C
–150 to 1200 °C
–100 to 1200 °C
–100 to 1300 °C
300 to 1760 °C
400 to 1760 °C
–100 to 400 °C
–200 to 600 °C
–80 to 150 °C
DC current (34901A only)
10.00000 mA
100.0000 mA
1.000000 A
True RMS AC current (34901A only)
+ 0.0020
+ 0.0005
+ 0.0010 + 0.006
+ 0.006
+ 0.006
+ 0.06
+ 0.06
10.00000 mA and 4 1.000000 A
100.0000 mA 8
B
E
J
K
N
R
S
T
R0 from 49 Ω to 2.1 kΩ
2.2 k, 5 k, 10 k
0.015 0.03 °C
+ 0.06
Temperature
Type
1-year accuracy 9
Extended range 1-year accuracy 9
Temp coefficient/oC
Thermocouple 10
RTD
Thermistor
0.003 °C
0.002 °C
1. Specifications are for 1 hr warm-up and 61⁄2 digits, Slow ac filter.
2. Relative to calibration standards.
3. 20% over range on all ranges except 300 Vdc and ac ranges and
2-wire ohms using scaling to remove error in 2-wire Ohms function.
current ranges.
4. For sinewave input > 5% of range. For inputs from 1% to 5% of range and < 50 kHz,
add 0.1% of range additional error.
5. Typically 30% of reading error at 1 MHz, limited to 1 x 108 V Hz.
6.
7. 8. 9. 10.
Specifications are for 4- wire ohms function or
the offset. Without scaling, add 4 Ω additional
Input > 100 mV. For 10 mV to 100 mV inputs multiply % of reading error x 10. Specified only for inputs > 10 mA.
For total measurement accuracy, add temperature probe error.
Thermocouple specifications not guaranteed when 34907A module is present. For < 1 °C accuracy, a precision external reference is required.
Find us at www.keysight.com
Page 16
Measurement characteristics 7
DC voltage
Measurement Method
A/D linearity Input resistance
100 mV, 1 V, 10 V ranges
100 V, 300 V ranges Input bias current Input protection
True RMS AC voltage
Measurement method
Crest factor Additional crest factor
errors (non-sinewave)
Input impedance Input protection
Resistance
Measurement method
Offset compensation Maximum lead resistance
Input protection
Frequency and period
Measurement method Voltage ranges
Gate time Measurement timeout
DC current
Shunt resistance Input protection
True RMS AC current
Measurement method
Shunt resistance Input protection
Thermocouple
Conversion
Reference junction type Open thermocouple check
Thermistor RTD
Measurement noise rejection 60 (50) Hz 1
Continuously Integrating
Multi-slope III A-D converter
0.0002% of reading + 0.0001 % of range
Selectable 10 MΩ or > 10,000 MΩ 10 MΩ ± 1%
< 30 pA at 25 °C
300 V all ranges
AC coupled True RMS — measures the AC component of the input with up to 300 Vdc of bias on any range
Maximum of 5:1 at Full Scale
dc CMRR
ac CMRR
Integration time
200 plc/3.33s (4s)
100 plc/1.67s (2s)
20 plc/333 ms (400 ms) 10 plc/167 ms (200 ms) 2 plc/33.3 ms (40 ms) 1 plc/16.7 ms (20 ms) < 1 plc
140 dB
70 dB
Normal mode rejection 2 110 dB 3
105 dB 3
100 dB 3
95 dB 3
90 dB
60 dB
0 dB
Crest Factor 1-2
Crest Factor 2-3
Crest Factor 3-4
Crest Factor 4-5
1 MΩ ± 2% in parallel with 150 pF 300 Vrms all ranges
Operating characteristics
Single channel measurement rates 5
0.05 % of reading 0.15 % of reading 0.30 % of reading 0.40 % of reading
Function
dcV, 2-wire resistance
Thermocouple
RTD, thermistor
acV
Frequency, period
Resolution 8
34970A/34972A readings/sec
6 (5) 54 (47) 500
6 (5) 52 (47) 280
6 (5) 49 (47) 200
4
61⁄2 digits (10 plc) 51⁄2 digits (1 plc) 41⁄2 digits (0.02 plc)
0.1 °C (10 plc) 0.1 °C (1 plc) (0.02 plc)
0.01 °C (10 plc) 0.1 °C (1 plc)
1 °C (0.02 plc)
61⁄2 Slow (3 Hz)
61⁄2 Med (20 Hz)
61⁄2 Fast (200 Hz)
61⁄26 100
61⁄2 digits (1 s gate) 1 51⁄2 digits (100 ms) 9 41⁄2 digits (10 ms) 70
Selectable 4-wire or 2-wire Ohms Current source referenced to LO input Selectable on 100 Ω, 1 kΩ, 10 kΩ ranges 10% of range per lead for 100 Ω and 1 kΩ ranges. 1 kΩ on all other ranges
300 V on all ranges
Reciprocal counting technique
Same as AC Voltage function
1 s, 100 ms, or 10 ms
Selectable 3 Hz, 20 Hz, 200 Hz LF limit
5Ω for 10 mA, 100 mA; 0.1 Ω for 1 A 1 A 250 V fuse on 34901A module
Direct coupled to the fuse and shunt.
AC coupled True RMS measurement (measures the ac component only)
5 Ω for 10 mA; 0.1 Ω for 100 mA, 1 A
1 A 250 V fuse on 34901A module
ITS-90 software compensation Internal, Fixed, or External Selectable per channel. Open > 5 kΩ
44004, 44007, 44006 series
α = 0.00385 (DIN) and α = 0.00391
0.14 1 8
1. For 1 KΩ unbalance in LO lead.
2. For power line frequency ± 0.1%.
3. For power line frequency ± 1% use 80 dB or ± 3% use 60 dB.
4. Reading speeds for 60 Hz and (50 Hz) operation.
5. For fixed function and range, readings to memory, scaling and alarms off,
AZERO OFF, USB datalogging OFF.
6. Maximum limit with de fault settling delays defeated.
7. Isolation voltage (ch-ch, ch-earth) 300 Vdc, ac rms.
8. 61⁄2 digits = 22 bits, 51⁄2 digits = 18 bits, 41⁄2 digits = 15 bits.
Find us at www.keysight.com
Page 17
Single channel reading rates to I/O or internal memory
(See footnotes for conditions below)
34970A
34972A
Into memory readings/sec
To GPIB or RS232 readings/sec
To LAN, USB or memory readings/sec
Single channel ASCII readings 500
Single channel while changing scale 25 (eg MEAS dcV 10/MEAS dcV 1)
Single channel while changing function 12 (eg MEAS dcV/MEAS Ohms)
Scanning measurement rates to I/O or internal memory
34901A/34908A 60
34902A 250
34902A into and out of memory (using INIT, FETCh) —
34902A with timestamp (using MEAS) —
34902A with scaling and alarms 220
34902A dcV and ohms on alternate channels 80 34901A/34908A 50
34902A 100 34901A/34908A 50
34902A 160 34907A Digital Input 275
34907A Totalizer 240
Data out of memory 3, 4
(FETCh of 50K readings)
Readings 800
Readings with timestamp 450
440
25
12
500
25
12
60
240
240
240
220
80 50
100 50
150 275
240
120 K 60 K
(See footnotes for conditions below)
34970A
34972A
Into memory ch/sec
To GPIB or RS232 ch/sec
To LAN, USB or memory ch/sec
Scanning dcV or ohms channels
60
210
180
150
190
80 50
90 50
150 250
210
Scanning acV channels 2
Scanning temperature – thermistor or TC channels
Scanning digital in/totalizer channels
(See footnotes for conditions below)
34970A
34972A
Single channel 1, 2
Over GPIB readings/sec
Over RS232 readings/sec
Over USB readings/sec
Over LAN or memory readings/sec
600
320
55 K
35 K
1. Speeds are for 4 1/2 digits, delay 0, display off, USB data logging off, autozero off, auto-ranging off unless otherwise noted. Use MEAS command for best I/O performance. RS232 at 115 Kbaud.
2. Maximum, with default delays defeated.
3. Assumes relative time format (time since start of scan).
4. Typical rates assuming lightly loaded PC and limited other traffic on I/Os. LAN rates assume use of socket connection; VXI11 will be less.